Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722...
Moderate severity
Unreviewed
Published
May 13, 2022
to the GitHub Advisory Database
•
Updated May 22, 2025
Description
Published by the National Vulnerability Database
Aug 13, 2018
Published to the GitHub Advisory Database
May 13, 2022
Last updated
May 22, 2025
Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G The models identified above, when paired with a remote controller and having the "easy bolus" and "remote bolus" options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.